Diagnostic performance of 2D shear wave elastography in predicting liver fibrosis in patients with chronic hepatitis
Mehmet Aksakal1, Suna Ozhan Oktar2, Halit Nahit Sendur2
1Department of Radiology, Gazi University, 06500, Beşevler, Ankara, Turkey. mehmetaksakalmd@gmail.com.
Purpose:
In this study, we investigated the diagnostic efficacy of 2D-Shear Wave Elastography (2D-SWE) in detecting stages of liver fibrosis and determining the disease-specific cut-off values in patients with chronic hepatitis B and C infection, using histopathological analysis as the reference method.
Method:
Our study included 103 consecutive adult patients with chronic hepatitis B and C infection (CHB and CHC) who had liver biopsy within three months of elastography examination. A real-time 2D-SWE was performed using the LOGIQ E9 system (GE Medical Systems, Wisconsin, USA). The correlation between the liver stiffness measurements and the METAVIR scores was evaluated. The diagnostic performance of 2D-SWE was assessed, and cut-off values were set.
Results:
We found a statistically significant positive correlation between elastography values and the degree of liver fibrosis (Spearman's correlation coefficient = 0.76 and 0.83 for CHB and CHC; respectively) (p = 0.0001). The stiffness cut-off values were F ≥ 1: 5.92 kPa, F ≥ 2: 7.69 kPa, F ≥ 3: 8.97 kPa, F ≥ 4: 12.15 kPa in CHB; and F ≥ 1: 6.09 kPa, F ≥ 2: 7.81 kPa, F ≥ 3: 9.0 kPa, F ≥ 4: 12.47 kPa in CHC patients.
Conclusion:
2D-SWE is reliable and accurate for the diagnosis of liver fibrosis. In selected patients, 2D-SWE may be useful in reducing the need for liver biopsy when staging fibrosis. Further studies in larger prospective series are needed to confirm these results and determine the most appropriate cut-off values for each stage of liver fibrosis.
More Related Videos
07:13Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test