Related Experiment Video
Updated: Nov 12, 2025

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Effectiveness of reaching and maintaining therapeutic hypothermia target temperature using low-cost devices in
Adriana Nishimoto Kinoshita1, Maurício Magalhães1,2,3,4, Daniela Pereira Rodrigues2,5
1Neonatal Division, Department of Pediatrics, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.
Abstract:
Therapeutic hypothermia (TH) using servo-controlled cooling devices has proved to be a safe and effective method to reduce mortality and sequelae in neonates with hypoxic-ischemic encephalopathy (HIE). However, such cooling devices can be expensive and have limited availability in low- and middle-income countries. To evaluate the feasibility and effectiveness of low-cost cooling devices (ice packs) to reach and maintain the target temperature in newborns with moderate or severe HIE. Descriptive retrospective cross-sectional study, including newborns with gestational age ≥35 weeks, weight ≥1800 g, with diagnosis of moderate or severe HIE, submitted to whole body hypothermia using ice packs for 72 hr. Data was obtained from medical records and databases. The thermal curves of each patient, time spent at the target temperature and rewarming time were evaluated. Complications directly related to therapeutic hypothermia, evolution, and early outcomes before hospital discharge were analyzed. Seventy-one newborns met the criteria for TH, and 61 completed the entire cooling period and were included in the final analysis. The average time to reach the target temperature was 1.50 ± 1.3 hr. The mean temperature during the cooling period was 33.5°C (±0.62). 82.32% of the measurements were within the target temperature range (from 33 to 34°C). The following adverse events were observed during TH: pulmonary hypertension 11.3%, arrhythmia 4.2%, and coagulopathy 26.7%. Whole body hypothermia using low-cost devices proved to be an effective method in the absence of availability of servo-controlled devices, representing a plausible option in restricted resource settings.
More Related Videos
Related Concept Videos
Decreased Body Temperature
Methods of reducing fever
Pharmacological Methods of Reducing Fever:
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...