Related Experiment Video
Updated: Nov 12, 2025

Tachycardia-Induced Cardiomyopathy As a Chronic Heart Failure Model in Swine
Published on: February 17, 2018
Sleep Rate Mode of Pacemaker-Dependent Patients with Sick Sinus Syndrome Increases Dipper Blood Pressure and Dipper
Tomoyuki Kabutoya1, Yasushi Imai1, Takafumi Okuyama1
1Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine.
Insights
Pacemaker settings can influence diurnal blood pressure (BP) and heart rate (HR) patterns. The sleep rate mode, compared to DDD mode, improved dipper BP and HR patterns in patients with sick sinus syndrome.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Dipper blood pressure (BP) and heart rate (HR) patterns are associated with better cardiovascular outcomes.
- The impact of pacemaker modes on diurnal BP and HR patterns is not well understood.
Purpose of the Study:
- To investigate how different pacemaker modes affect diurnal blood pressure and heart rate patterns in patients with sick sinus syndrome.
Main Methods:
- Nine patients with sick sinus syndrome underwent sequential 6-month pacing regimens in DDD mode and sleep rate mode (randomized crossover).
- Ambulatory BP monitoring was conducted, and BP/HR patterns were classified (dipper, nondipper).
Main Results:
- Sleep rate mode significantly increased the proportion of patients with dipper HR patterns compared to baseline and DDD mode.
- Sleep rate mode also significantly increased dipper BP patterns compared to baseline, with a marginal difference compared to DDD mode.
- Nighttime HR was significantly lower in sleep rate mode than in DDD mode.
Conclusions:
- Pacemaker settings utilizing a sleep rate mode can enhance diurnal heart rate and blood pressure variability in pacemaker-dependent patients.
- This suggests potential benefits for cardiovascular health in patients with sick sinus syndrome managed with specific pacing strategies.
Abstract:
Cardiovascular event rates of patients with a dipper blood pressure (BP) and dipper heart rate (HR) pattern are lower than those of patients with nondipper BP and HR patterns. However, how the pacemaker mode affects the diurnal BP and HR patterns remains unclear.We enrolled nine patients (average age 74.4 ± 6.6 years, 4 males and 5 females) with sick sinus syndrome who required atrial pacing. We investigated sequential 6-month pacing regimens (DDD mode at 60 bpm and sleep rate mode). We set the lower rate of sleep rate mode as follows: 60 bpm during the daytime and 50 bpm during the nighttime. The order of pacing mode was randomized, with crossover design. Ambulatory BP monitoring was performed at baseline, 6 months, and 12 months, BP category was classified into four groups (extreme dipper, dipper, nondipper, and riser pattern), and HR was classified into dipper and nondipper patterns.Nighttime HR during the sleep rate mode was significantly lower than that at DDD (57.1 ± 6.2 versus 63.5 ± 3.8 bpm, P = 0.001). The dipper HR pattern was increased in the sleep rate mode compared with those at baseline or DDD mode (versus baseline: 89% versus 44%, P = 0.035; versus DDD: 89% versus 22%, P = 0.004). The dipper BP pattern significantly increased in the sleep rate mode compared with the baseline (56% versus 11%, P = 0.035), but the difference between the sleep rate mode and DDD mode was statistically marginal (56% versus 22%, P = 0.081).The pacemaker settings in the sleep rate mode increased the dipper HR and BP patterns in pacemaker-dependent patients with sick sinus syndrome.
More Related Videos
09:56Implantation of Combined Telemetric ECG and Blood Pressure Transmitters to Determine Spontaneous Baroreflex Sensitivity in Conscious Mice
Published on: February 14, 2021
09:20Microelectrode Array Recording of Sinoatrial Node Firing Rate to Identify Intrinsic Cardiac Pacemaking Defects in Mice
Published on: July 5, 2021
Related Concept Videos
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Regulation of Pulse