Liquid chromatography-tandem mass spectrometry method for measuring vitamin E acetate in bronchoalveolar lavage fluid
Maria Morel Espinosa1, Benjamin C Blount1, Liza Valentin-Blasini1
1Division of Laboratory Sciences, National Center for Environmental Health, U.S. Centers for Disease Control and Prevention, Atlanta, GA 30341, USA.
Abstract:
We investigated the suitability of isotope-dilution liquid chromatography coupled with tandem mass spectrometry for identifying vitamin E acetate (VEA) in bronchoalveolar lavage (BAL) fluid. This new method demonstrates high accuracy, selectivity, and sensitivity, with mean recoveries higher than 90%, coefficients of variation ranging from 1.5% to 4.5%, and a limit of detection of 1.10 ng/mL. Calibration curves were linear (R2 > 0.99). The linear range and detection limit of the method were adequate for identifying VEA in 48 of 51 BAL fluid samples collected from people with lung injury resulting from e-cigarettes, or vaping, product use. We conclude that this method is an effective tool for studying VEA accumulation in lungs caused by using e-cigarettes, or vaping, products that contain VEA.