Related Experiment Video
Updated: Nov 12, 2025

Piezo High Accuracy Surgical Osteal Removal PHASOR: A Technique for Improved Cranial Window Surgery in Mice
Published on: March 2, 2018
OUR EXPERIENCE WITH ACTIVE SENTRY AND CENTURION OZIL HANDPIECES
Aim:
To compare parameters of phacoemulsification using handpiece Active Sentry and Centurion Ozil.
Methods:
We have evaluated results of 200 eyes of 129 patients that were operated for cataract at the Department of Ophthalmology. All surgeries were performed by one surgeon (NJ). In 100 eyes handpiece Active Sentry was used and in 100 eyes handpiece Centurion Ozil was used. The intraocular pressure during surgery (IOP), cumulative dissipated energy (CDE), ultrasound time (U/S time) and estimated consumption of balanced salt solution (BSS) were evaluated.
Results:
Using handpiece Ozil Centurion the IOP was 65 mmHg, use handpiece Active Sentry enabled decrease safely peroperative IOP to 46 mm Hg withouth increase of fluctuation or declension of stability of the anterior chamber. The mean CDE a U/S time were significantly statistically decreased using Active Sentry versus Centurion Ozil handpieces. Difference in estimated consumption of balanced salt solution was not statistically significant.
Conclusion:
Our results proved that using handpiece Active Sentry enabled statistically significantly decrease IOP during phacoemulsification and eliminated post-occlusion surge. This brings several advantages: more painless surgery for patients and more user-friendly procedure for surgeon especially in challenging cases (intraoperative floppy iris syndrome or severe myopia).
More Related Videos
07:06Endaural Endoscopic Atticoantrotomy Retrograde Mastoidectomy using a Constant Suction Bone-drilling Technique
Published on: May 23, 2021
09:07Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
Related Concept Videos
Hand hygiene
Hand washing...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Suctioning II: Procedure