Related Experiment Video
Updated: Nov 12, 2025

Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Assessment of an Automatic Hot Water Sterilizing System-Equipped Dialysis Supply Device's Circuit Degradation, 7
Tomonari Ogawa1, Yuki Kanayama2, Yusuke Sasaki2
1From the Department of Nephrology and Blood Purification, Saitama Medical Center, Saitama Medical University, Kawagoe, Japan.
Abstract:
At most dialysis clinics in Japan, a central dialysis fluid delivery system (CDDS) is used primarily to provide consistent treatment to patients. We incorporated hot water disinfection to a CDDS over a 7 year period, after which we assessed the dialysis purification levels. We observed that adequate levels had been maintained. We rated the internal circuits of the device at the 7 year mark, comparing the insides of both the used and unused new Synflex tubing, silicone blades, Teflon tubing, silicone rubber tubing, and stainless steel parts. No bacteria or contamination (such as endotoxins) was detected, and no degradation or damage were observed. Even after the auto-hot water disinfection system was installed, no internal tubing degradation due to hot water or mechanical issues have occurred, and the system's dialysis fluid purification levels have been maintained.
Related Concept Videos
Continuous Renal Replacement Therapy
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Hemodialysis III: Nursing Management
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis II: Procedure and Complications
Hemodialysis I: Introduction