Considerations on Performance Evaluation of Atrial Fibrillation Detectors
Objective:
A large number of atrial fibrillation (AF) detectors have been published in recent years, signifying that the comparison of detector performance plays a central role, though not always consistent. The aim of this study is to shed needed light on aspects crucial to the evaluation of detection performance.
Methods:
Three types of AF detector, using either information on rhythm, rhythm and morphology, or segments of ECG samples, are implemented and studied on both real and simulated ECG signals. The properties of different performance measures are investigated, for example, in relation to dataset imbalance.
Results:
The results show that performance can differ considerably depending on the way detector output is compared to database annotations, i.e., beat-to-beat, segment-to-segment, or episode-to-episode comparison. Moreover, depending on the type of detector, the results substantiate that physiological and technical factors, e.g., changes in ECG morphology, rate of atrial premature beats, and noise level, can have a considerable influence on performance.
Conclusion:
The present study demonstrates overall strengths and weaknesses of different types of detector, highlights challenges in AF detection, and proposes five recommendations on how to handle data and characterize performance.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...