Aims:
This work aims at presenting a fully coupled approach for the numerical solution of contact problems between multiple elastic structures immersed in a fluid flow. The key features of the computational model are (i) a fully coupled fluid-structure interaction with contact, (ii) the use of a fibre-reinforced material for the leaflets, (iii) a stent, and (iv) a compliant aortic root.
Methods And Results:
The computational model takes inspiration from the immersed boundary techniques and allows the numerical simulation of the blood-tissue interaction of bioprosthetic heart valves (BHVs) as well as the contact among the leaflets. First, we present pure mechanical simulations, where blood is neglected, to assess the performance of different material properties and valve designs. Secondly, fully coupled fluid-structure interaction simulations are employed to analyse the combination of haemodynamic and mechanical characteristics. The isotropic leaflet tissue experiences high-stress values compared to the fibre-reinforced material model. Moreover, elongated leaflets show a stress concentration close to the base of the stent. We observe a fully developed flow at the systolic stage of the heartbeat. On the other hand, flow recirculation appears along the aortic wall during diastole.
Conclusion:
The presented FSI approach can be used for analysing the mechanical and haemodynamic performance of a BHV. Our study suggests that stresses concentrate in the regions where leaflets are attached to the stent and in the portion of the aortic root where the BHV is placed. The results from this study may inspire new BHV designs that can provide a better stress distribution.
Related Concept Videos
Fluid Pressure over Curved Plate of Constant Width
Pressure of Fluids
Steady, Laminar Flow Between Parallel Plates
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Fluid Pressure over Flat Plate of Constant Width
The resultant force...
Navier–Stokes Equations