Discrepancies in dose and dose specification in interstitial implants.

J M Burgers1, H K Awwad, R van der Laarse

  • 1Department of Radiotherapy, The Netherlands Cancer Institute (Antoni von Leeuwenhoekhuis), Amsterdam.

Radiotherapy and Oncology : Journal of the European Society for Therapeutic Radiology and Oncology
|April 1, 1988
PubMed
Summary

Interstitial implant systems like Paterson-Parker and Paris offer guidance for source arrangement and dose calculation. However, variations in these systems can lead to significant dose rate discrepancies, impacting treatment accuracy.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Design and characterization of a new high-dose-rate brachytherapy Valencia applicator for larger skin lesions.

Medical physics·2016
Same author

Scleroma of the upper air-passages; a clinico-radiological study of 84 cases.

The Journal of the Faculty of Radiologists. Faculty of Radiologists (Great Britain)·2014
Same author

Dosimetric characterization of Ir-192 LDR elongated sources.

Medical physics·2008
Same author

Design and evaluation of a HDR skin applicator with flattening filter.

Medical physics·2008
Same author

Technique for routine output verification of Leipzig applicators with a well chamber.

Medical physics·2006
Same author

Comment on 'Comparison of dose rates calculated on Nucletron NPS v11: catheter tracking versus catheter describing'.

Physics in medicine and biology·2003

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Brachytherapy

Background:

  • Interstitial implant brachytherapy relies on established systems (e.g., Paterson-Parker, Paris) for source arrangement and dose calculation.
  • These systems aim to ensure adequate coverage of the planned target volume and define reference dose rates.

Purpose of the Study:

  • To investigate dose specification issues in interstitial implants.
  • To compare predicted reference dose rates with computer-calculated doses for theoretical implants.
  • To quantify discrepancies arising from deviations from standard implant system rules.

Main Methods:

  • Theoretical interstitial implants were analyzed.
  • Predicted reference dose rates were compared against computer-calculated dose rates for the planned target volume.

Related Experiment Videos

  • Deviations from standard implant system rules were systematically introduced and evaluated.
  • Main Results:

    • Discrepancies between predicted and calculated dose rates were observed.
    • These discrepancies increased significantly with moderate deviations from established implant system rules, common in clinical practice.
    • Specific examples illustrate the magnitude and location of dose rate changes exceeding 10%.

    Conclusions:

    • Clinicians must be aware of potential dose variations when using interstitial implant systems.
    • Adherence to established rules is crucial for minimizing dose discrepancies and ensuring treatment efficacy.
    • Understanding these variations is essential for optimizing interstitial brachytherapy outcomes.