Commissioning cranial single-isocenter multi-target radiosurgery for the Versa HD
Cory Knill1, Raminder Sandhu1, Robert Halford1
1Department of Radiation Oncology, Beaumont Health, Royal Oak, MI, 48073, USA.
Journal of Applied Clinical Medical Physics
|March 23, 2021
Summary
Commissioning Brainlab
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Single-isocenter multi-target (SIMT) cranial stereotactic radiosurgery (SRS) requires specialized treatment planning.
- Brainlab's Elements Multiple Brain Mets SRS (MBMS) is designed for these complex treatments.
Purpose of the Study:
- To document the commissioning process of the MBMS system on an Elekta Versa HD linear accelerator.
- To evaluate the accuracy and safety of MBMS for cranial SRS.
Main Methods:
- Commissioning involved collecting beam data (output factors, PDDs, profiles) for 6 MV, 6 MV FFF, and 10 MV FFF photon beams.
- Treatment plans were created for patient data and evaluated using conformity and gradient indices.
- Dosimetric accuracy was verified using microDiamond, Gafchromic film, and SRS Mapcheck, with end-to-end testing for the entire workflow.
Main Results:
- Beam model parameters were within acceptable tolerances.
- MBMS plans achieved excellent conformity (ICI < 1.5) and gradient indices (GI < 5.0) for targets > 10 mm.
- Measurements confirmed high dosimetric accuracy, with end-to-end tests showing dose agreement within 1.5%.
Conclusions:
- The commissioning of Brainlab's MBMS on the Elekta Versa HD was successful.
- This experience provides a valuable guide for implementing safe and accurate SIMT cranial SRS treatments.