Validity and reproducibility of foot motion analysis using a stretch strain sensor.
Kodai Sakamoto1, Chie Tsujioka2, Megumi Sasaki2
1Graduate School of Health Science, Morinomiya University of Medical Sciences, 1-26-16 Nankokita Suminoe Ward Osaka City Osaka Prefecture, 559-8611, Japan.
A new stretch strain sensor (STR) shows good accuracy for analyzing foot motion, even with shoes on. This portable technology offers a reliable method for clinical and real-world gait analysis.
Area of Science:
- Biomechanics
- Sports Medicine
- Orthopedics
Background:
- Multi-segment foot analysis is challenging with footwear or orthoses in clinical settings.
- A novel, thin, and flexible stretch strain sensor (STR) offers a portable solution without camera observation.
Purpose of the Study:
- To evaluate the reproducibility and validity of foot motion analysis using the STR.
- To compare STR data with a conventional motion capture system during walking and running.
Main Methods:
- Twenty-one healthy participants underwent simultaneous kinematic data collection using STR and 3D motion capture.
- STR was applied at the spring ligament (SL) and navicular drop (ND) sites.
- Reproducibility assessed via Intra-class correlation (ICC); validity via cross-correlation and Pearson/Spearman correlation.
Main Results:
- The SL method demonstrated high ICCs (0.95-0.96) and strong correlation with forefoot frontal motion (r=-0.64 to -0.67).
- STR measurements significantly correlated with forefoot eversion excursion.
- The ND method showed less association with forefoot and hindfoot kinematics (ICCs 0.93 and 0.71).
Conclusions:
- The STR system exhibits acceptable reproducibility and validity for foot motion analysis.
- This technology facilitates in-shoe foot motion measurement, expanding applications beyond laboratory settings.