Mean Absolute Relative Difference of Blood Glucose Monitoring Systems and Relationship to ISO 15197
Guido Freckmann1, Jochen Mende1, Stefan Pleus1
1Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Background:
The analytical quality of a blood glucose monitoring system (BGMS) is often assessed according to the requirements described in the international standard ISO 15197. However, the mean absolute relative difference (MARD) is sometimes used as well. This analysis aims at providing empirical data from BGMS evaluation studies conducted according to ISO 15197 and at providing an estimation of how MARD and percentage of measurement results within ISO accuracy limits are related.
Methods:
Results of 77 system accuracy evaluations conducted according to ISO 15197 were used to calculate MARD between BGMS and a laboratory comparison method's results (glucose oxidase or hexokinase method). Additionally, bias and 95%-limits of agreement (LoA) using the Bland and Altman method were calculated.
Results:
MARD results ranged from 2.3% to 20.5%. The lowest MARD of a test strip lot that showed <95% of results within ISO limits was 6.1%. The distribution of MARD results shows that only 3.6% of test strip lots with a MARD equal to or below 7% showed <95% of results within ISO limits (2.2% of all test strip lots). Bias of test strip lots that showed ≥95% of results within the limits ranged from -10.3% to +7.4%. The half-width of the 95%-LoA of test strip lots that showed ≥95% of results within the limits ranged from 4.8% to 24.0%.
Conclusion:
There is a threshold MARD that may allow an estimate whether ISO 15197 requirements are fulfilled, but this statement cannot be made with certainty.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Bioequivalence Data: Statistical Interpretation
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...