Platelet activation and aggregation in different centrifugal-flow left ventricular assist devices
Maximilian Tscharre1,2, Franziska Wittmann3, Daniela Kitzmantl2
1Department of Internal Medicine, Cardiology and Nephrology, Landesklinikum Wiener Neustadt, Wiener Neustadt, Austria.
Platelets
|March 24, 2021
Summary
Platelet activation and aggregation markers were similar between HeartMate 3 (HM3) and Medtronic/Heartware HVAD (HVAD) left-ventricular assist devices (LVADs). However, HVAD patients showed lower von Willebrand factor activity and fewer large VWF multimers.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Hematology
Background:
- Left-ventricular assist devices (LVADs) are crucial for end-stage heart failure management.
- Two centrifugal-flow LVADs, HeartMate 3 (HM3) and Medtronic/Heartware HVAD (HVAD), are approved.
- Differences in thrombogenicity between HM3 and HVAD warrant investigation.
Purpose of the Study:
- To compare platelet activation and aggregation markers between HM3 and HVAD systems.
- To investigate potential differences in von Willebrand factor (VWF) profiles between the two LVADs.
Main Methods:
- Prospective study of 59 LVAD patients (40 HM3, 19 HVAD).
- Flow cytometry assessed P-selectin, GPIIb/IIIa, and monocyte-platelet aggregates (MPA).
- Light-transmission aggregometry (LTA) and multiple-electrode aggregometry (MEA) measured platelet aggregation.
- VWF antigen (VWF:Ag), VWF activity (VWF:Ac), and VWF multimer patterns were analyzed.
- Soluble P-selectin (sP-selectin) was measured via ELISA.
Main Results:
- In vivo and agonist-stimulated platelet activation markers (P-selectin, GPIIb/IIIa, MPA) were similar between HM3 and HVAD.
- Platelet aggregation measured by LTA and MEA did not differ significantly between the two LVAD groups.
- VWF:Ag and FVIII:C levels were comparable, but HVAD patients exhibited significantly lower VWF:Ac and reduced large VWF multimers.
- Soluble P-selectin levels were similar in both LVAD groups.
Conclusions:
- On-treatment platelet activation and aggregation are comparable between HeartMate 3 and HVAD left-ventricular assist device patients.
- Observed differences in von Willebrand factor activity and multimer profiles between HVAD and HM3 require further clinical investigation.
- Future trials should explore the clinical implications of these VWF differences in LVAD thrombogenicity.