Analysis and mitigation of noise in simultaneous transmission and reception in MRI
Bilal Tasdelen1,2, Alireza Sadeghi-Tarakameh1,2, Ugur Yilmaz2
1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey.
Purpose:
In simultaneous transmission and reception (STAR) MRI, along with the coupling of the excitation pulse to the received signal, noise, and undesired distortions (spurs) coming from the transmit chain also leak into the acquired signal and degrade image quality. Here, properties of this coupled noise and its relationship with the transmit amplifier gain, transmit chain noise density, isolation performance, and imaging bandwidth are analyzed. It is demonstrated that by utilizing a recently proposed STAR technique, the transmit noise can be reduced. The importance of achieving high isolation and careful selection of the corresponding parameters are demonstrated.
Theory And Methods:
A cancellation algorithm, together with a vector modulator, is used for transmit-receive isolation. The scanner is modeled as a pipeline of blocks to demonstrate the noise contribution from each block. With higher isolation, coupled transmit noise can be reduced to the point that the dominant noise source becomes acquisition noise, as in the case for pulsed MRI. Amplifiers with different gain and noise properties are used in the experiments to verify the derived noise-transmit parameter relation.
Results:
With the proposed technique, more than 80 dB isolation in the analog domain is achieved. The leakage noise and the spurs coupled from the transmit chain, are reduced. It is shown that the transmit gain plays the most critical role in determining sufficient isolation, whereas the amplifier noise figure does not contribute as much.
Conclusion:
The transmit noise and the spurs in STAR imaging are analyzed and mitigated by using a vector modulator.
More Related Videos
11:28Concurrent EEG and Functional MRI Recording and Integration Analysis for Dynamic Cortical Activity Imaging
Published on: June 30, 2018
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...