Repeatability of 2 swept-source OCT biometers and 1 optical low-coherence reflectometry biometer
Andreea D Fişuş1, Nino D Hirnschall, Manuel Ruiss
1From the Department of Ophthalmology, Hanusch Hospital, Vienna Institute for Research in Ocular Surgery-Karl Landsteiner Institute, Vienna, Austria .
Journal of Cataract and Refractive Surgery
|March 26, 2021
Summary
Swept-source optical coherence tomography (SS-OCT) biometers demonstrated superior repeatability compared to optical low-coherence reflectometry (OLCR) for cataract surgery measurements. Both SS-OCT devices showed high precision in keratometry, anterior chamber depth, and axial length.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Devices
Background:
- Accurate ocular biometry is crucial for successful cataract surgery outcomes.
- Advancements in biometry devices, including swept-source optical coherence tomography (SS-OCT) and optical low-coherence reflectometry (OLCR), aim to improve measurement precision.
- Comparing the repeatability of these technologies is essential for clinical adoption.
Purpose of the Study:
- To compare the repeatability of two SS-OCT biometers (IOLMaster 700 and ANTERION) and one OLCR device (LENSTAR LS900).
- To evaluate the precision of key biometric parameters: keratometry, central corneal thickness, anterior chamber depth (ACD), lens thickness (LT), and axial eye length (AL).
Main Methods:
- A prospective study involving 50 patients scheduled for cataract surgery.
- Three consecutive measurements were taken with each of the two SS-OCT devices and the OLCR device.
- Repeatability was assessed using within-subject standard deviation (Sw) and coefficient of variation (CoV).
Main Results:
- All devices exhibited high repeatability, with CoV values below 0.5 for most parameters.
- SS-OCT devices generally showed lower Sw values for keratometry, ACD, and AL compared to the OLCR device.
- The ANTERION (SS-OCT) demonstrated the highest repeatability for ACD (Sw=0.004).
Conclusions:
- Both SS-OCT biometry devices offer excellent repeatability for ocular measurements.
- SS-OCT technology demonstrates a higher repeatability performance than OLCR for key biometry variables.
- These findings support the clinical utility of SS-OCT devices in optimizing cataract surgery planning.