Computed Tomography Perfusion Analysis of Pancreatic Adenocarcinoma using Deconvolution, Maximum Slope,
Marco Koell1, Miriam Klauss1, Stephan Skornitzke1
1Clinic of Diagnostic and Interventional Radiology, University of Heidelberg, Germany.
Purpose:
The goal of this study was to evaluate the diagnostic accuracy of perfusion computed tomography (CT) parameters obtained by different mathematical-kinetic methods for distinguishing pancreatic adenocarcinoma from normal tissue. To determine cut-off values and to assess the interchangeability of cut-off values, which were determined by different methods.
Materials And Methods:
Perfusion CT imaging of the pancreas was prospectively performed in 23 patients. 19 patients with histopathologically confirmed pancreatic adenocarcinoma were included in the study. Blood flow (BF), blood volume (BV) and permeability-surface area product (PS) were measured in pancreatic adenocarcinoma and normal tissue with the deconvolution (BF, BV, PS), maximum slope (BF), and Patlak methods (BV, PS). The interchangeability of cut-off values was examined by assessing agreement between BF, BV, and PS measured with different mathematical-kinetic methods.
Results:
Bland-Altman analysis demonstrated poor agreement between perfusion parameters, measured with different mathematical-kinetic methods. According to receiver operating characteristic (ROC) analysis, PS measured with the Patlak method had the significantly lowest diagnostic accuracy (area under ROC curve = 0.748). All other parameters were of high diagnostic accuracy (area under ROC curve = 0.940-0.997), although differences in diagnostic accuracy were not statistically different. Cut-off values for BF of ≤ 91.83 ml/100 ml/min and for BV of ≤ 5.36 ml/100 ml, both measured with the deconvolution method, appear to be the most appropriate cut-off values to distinguish pancreatic adenocarcinoma from normal tissue.
Conclusion:
Perfusion parameters obtained by different methods are not interchangeable. Therefore, cut-off values, which were determined using different methods, are not interchangeable either. Perfusion parameters can help to distinguish pancreatic adenocarcinoma from normal tissue with high diagnostic accuracy, except for PS measured with the Patlak method.
Key Points:
· Perfusion CT parameters showed high diagnostic accuracy in differentiating between pancreatic adenocarcinoma and normal tissue.. · Only PS measured with the Patlak method showed a significantly lower diagnostic accuracy.. · Perfusion parameters measured with different mathematical-kinetic methods are not interchangeable.. · A specific cut-off value must be determined for each method and each perfusion parameter..
Citation Format:
· Koell M, Klauss M, Skornitzke S et al. Computed Tomography Perfusion Analysis of Pancreatic Adenocarcinoma with the Deconvolution, Maximum Slope, and Patlak Methods - Evaluation of Diagnostic Accuracy and Interchangeability of Cut-Off Values. Fortschr Röntgenstr 2021; 193: 1062 - 1073.
More Related Videos
06:24Dynamic Contrast Enhanced Magnetic Resonance Imaging of an Orthotopic Pancreatic Cancer Mouse Model
Published on: April 18, 2015
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT