Related Experiment Video
Updated: Nov 11, 2025

Investigating von Willebrand Factor Pathophysiology Using a Flow Chamber Model of von Willebrand Factor-platelet String Formation
Published on: August 14, 2017
Evaluation of a new semi-automated Hydragel 11 von Willebrand factor multimers assay kit for routine use
Marika Pikta1, Timea Szanto2, Margus Viigimaa3
1North Estonia Medical Centre, Department of Laboratory Medicine, Tallinn, Estonia.
Background:
Accurate diagnosis and classification of von Willebrand disease (VWD) are essential for optimal management. The von Willebrand factor multimers analysis (VWF:MM) is an integral part of the diagnostic process in the phenotypic classification, especially in discrepant cases. The aim of this study was to evaluate the performance of a new Hydragel 11VWF multimer assay (H11VW).
Methods:
Analytical performance characteristics such as repeatability (intra-assay variability, in gel between track variation), reproducibility (inter-assay variability, between gel variation), sensitivity, EQA performance and differences between two commercially available VWF:MM kits (H5VW and H11VW) were analysed in healthy volunteers' plasmas using in-house prepared reference plasma.
Results:
Repeatability and reproducibility results of H11VW demonstrated acceptable and equivalent performance with previously verified H5VW. Participation in EQA was successful. No statistically significant difference was detected between H5VW and H11VW kits for different fractions of multimers: LMWM p=0.807; IMWM p=0.183; HMWM p=0.774.
Conclusions:
H11VW demonstrated acceptable analytical performance characteristics. H11VW kit conveniently offers a more significant number of samples on a single gel. H5VW and H11VW kits can be used in daily practice interchangeably.