Intravenous Line Phase-Wrap Artifact at Bilateral Axial 3-T Breast MRI: Identification, Analysis, and Solution
Einat Slonimsky1, Yusef Azraq1, John M Gomori1
1Department of Diagnostic Imaging, Department of Radiology, Penn State Health Milton S Hershey Medical Center, Penn State University Hospital, 500 University Dr, Hershey, PA 17033-0850 (E.S.); Department of Diagnostic Imaging, Hadassah Hebrew University Medical Center, Jerusalem, Israel, Affiliated with the Hebrew University Medical School, Jerusalem, Israel (Y.A., J.M.G., S.F., T.S.); and Department of Diagnostic Imaging, Edith Wolfson Medical Center, Holon, Israel, Affiliated with the Sackler School of Medicine, Tel Aviv University, Tel Aviv, Israel (T.A.K.).
Purpose:
To understand and remove the source of a phase-wrap artifact produced by residual contrast agent in the intravenous line during acquisition of bilateral axial 3-T dynamic contrast material-enhanced (DCE) breast MRI.
Materials And Methods:
A two-part study involved a phantom experiment, followed by an institutional review board approved clinical intervention, to evaluate the phase-wrap artifact at MRI. A phantom model evaluated artifact production by using an intravenous line filled with fluids with varying concentrations of gadolinium-based contrast agent (0, 0.4, 0.8, 1.2, 1.6, and 2 mmol/mL) and by positioning the simulated intravenous line within several fields of view (FOV) at 3-T MRI in breast coils. Next, a clinical assessment was performed with a total of 400 patients (control group:interventional group, 200:200) to determine the effect of taping the intravenous line to the patients' backs. Breast MR images were assessed blindly for the presence of the artifact. Software was used for statistical analysis with a P value of less than .05 considered a significant difference.
Results:
In the phantom model, the artifact was produced only with a 0.4 mmol/mL gadolinium concentration and when the tubing was either close to the edge or within a FOV of 350-450 mm. In the clinical experiment, the artifact was more prevalent in the retrospective control group than in the prospective intervention group (52.5% [105 of 200] vs 22% [44 of 200]; P < .005).
Conclusion:
The presence of phase-wrap artifacts can be reduced by moving the contrast agent intravenous line out of the FOV during acquisition by taping it to a patient's back during bilateral axial 3-T DCE breast MRI.Keywords: Breast, MR-Imaging, Phantom Studies© RSNA, 2020.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
10:26A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
Published on: June 2, 2015
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Magnetic Resonance Imaging