Related Experiment Video
Updated: Nov 11, 2025

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Smartphone electrocardiogram for detecting atrial fibrillation after a cerebral ischaemic event: a multicentre
Keng Tat Koh1, Wan Chung Law2, Win Moe Zaw2
1Department of Cardiology, Sarawak Heart Centre, 3rd Roundabout, Samarahan Expressway, 94300 Kota Samarahan, Sarawak, Malaysia.
Aims:
Atrial fibrillation (AF) is a preventable cause of ischaemic stroke but it is often undiagnosed and undertreated. The utility of smartphone electrocardiogram (ECG) for the detection of AF after ischaemic stroke is unknown. The aim of this study is to determine the diagnostic yield of 30-day smartphone ECG recording compared with 24-h Holter monitoring for detecting AF ≥30 s.
Methods And Results:
In this multicentre, open-label study, we randomly assigned 203 participants to undergo one additional 24-h Holter monitoring (control group, n = 98) vs. 30-day smartphone ECG monitoring (intervention group, n = 105) using KardiaMobile (AliveCor®, Mountain View, CA, USA). Major inclusion criteria included age ≥55 years old, without known AF, and ischaemic stroke or transient ischaemic attack (TIA) within the preceding 12 months. Baseline characteristics were similar between the two groups. The index event was ischaemic stroke in 88.5% in the intervention group and 88.8% in the control group (P = 0.852). AF lasting ≥30 s was detected in 10 of 105 patients in the intervention group and 2 of 98 patients in the control group (9.5% vs. 2.0%; absolute difference 7.5%; P = 0.024). The number needed to screen to detect one AF was 13. After the 30-day smartphone monitoring, there was a significantly higher proportion of patients on oral anticoagulation therapy at 3 months compared with baseline in the intervention group (9.5% vs. 0%, P = 0.002).
Conclusions:
Among patients ≥55 years of age with a recent cryptogenic stroke or TIA, 30-day smartphone ECG recording significantly improved the detection of AF when compared with the standard repeat 24-h Holter monitoring.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Acute Coronary Syndrome III: Diagnostic Studies
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias