Fire safety study on high-flow nasal oxygen in shared-airway surgeries with diathermy and laser: simulation based on
Man-Yun Chang1, Jui-Hung Chen1, Shih-Pin Lin1
1Department of Anesthesiology, Taipei Veterans General Hospital, No. 201, Shih-Pai Rd. 2nd, Beitou, Taipei City, Taiwan.
Abstract:
High-flow nasal oxygen (HFNO) has been used in "tubeless" shared-airway surgeries but whether HFNO increased the fire hazard is yet to be examined. We used a physical model for simulation to explore fire safety through a series of ignition trials. An HFNO device was attached to a 3D-printed nose with nostrils connected to a degutted raw chicken. The HFNO device was set at twenty combinations of different oxygen concentration and gas flow rate. An electrocautery and diode laser were applied separately to a fat cube in the cavity of the chicken. Ten 30 s trials of continuous energy source application were conducted. An additional trial of continuous energy application was conducted if no ignition was observed for all the ten trials. A total of eight short flashes were observed in one hundred electrocautery tests; however, no continuous fire was observed among them. There were thirty-six events of ignition in one hundred trials with laser, twelve of which turned into violent self-sustained fires. The factors found to be related to a significantly increased chance of ignition included laser application, lower gas flow, and higher FiO2. The native tissue and smoke can ignite and turn into violent self-sustained fires under HFNO and continuous laser strikes, even in the absence of combustible materials. The results suggest that airway surgeries must be performed safely with HFNO if only a short intermittent laser is used in low FiO2.
More Related Videos
05:43Topical Airway Anesthesia for Awake-endoscopic Intubation Using the Spray-as-you-go Technique with High Oxygen Flow
Published on: January 13, 2017
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...