Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Implementation of an enclosed ionization interface for the analysis of liquid sample streams with direct analysis in
Ricarda Kloth1,2, Peyman Khanipour1,2, Karl J J Mayrhofer1,2
1Helmholtz Institute Erlangen-Nürnberg for Renewable Energy (IEK-11), Forschungszentrum Jülich GmbH, Egerlandstr. 3, Erlangen, 91058, Germany.
Rationale:
The development of an interface to analyze liquid sample streams with direct analysis in real time mass spectrometry (DART-MS) is of great interest for coupling various analytical techniques, using non-volatile salts, with MS. Therefore, we devised an enclosed ionization interface and a sample introduction system for the versatile analysis of liquid samples with DART-MS.
Methods:
The sample introduction system consists of a nebulizer, a spray chamber and a transfer line, while the confined ionization interface is created by implementing a cross-shaped housing between ion source outlet and mass spectrometer inlet. Methodical studies of the effects of various setup parameters on signal intensity and peak shape were conducted, while its diverse applicability was demonstrated by coupling with high-performance liquid chromatography (HPLC) for the analysis of alcohols, organic acids and furanic compounds.
Results:
The confinement of the ionization interface results in a robust setup design with a well-defined ionization region for focusing of the sprayed sample mist. Thereby, an increase in analyte signal intensity by three orders of magnitude and improved signal stability and reproducibility were obtained in comparison with a similar open ionization interface configuration. Additionally, the successful quantification of alcohols could be demonstrated as well as the compatibility of the setup with HPLC gradient elution.
Conclusions:
A versatile setup design for the analysis of liquid sample streams with DART-MS was devised for monitoring reactions or hyphenating analytics with MS. The design minimizes interferences from the laboratory surroundings as well as allows for safe handling of hazardous and toxic chemicals, which renders it suitable for a broad range of applications.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry
Mass Spectrometers
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Ion-Exchange Chromatography