Desflurane consumption with automated vapour control systems in two different anaesthesia machines. A randomized
Dagmar Mostad1, Pål Klepstad1,2, Turid Follestad3
1Clinic of Anaesthesia and Intensive Care, St. Olav's University Hospital, Trondheim, Norway.
Automated gas delivery devices significantly reduce desflurane consumption in anesthesia compared to conventional low flow methods. Both automated systems, Flow-I and Aisys, proved reliable for reducing volatile anesthetic use.
Area of Science:
- Anesthesiology
- Medical Devices
- Pharmacoeconomics
Background:
- General anesthesia typically uses a fresh gas flow (FGF) of at least 0.5 L/min.
- Automated gas delivery devices are designed to minimize volatile anesthetic consumption by controlling gas flow.
- This study investigates desflurane use with automated versus conventional low-flow anesthesia systems.
Purpose of the Study:
- To compare desflurane consumption between automated gas control and conventional low-flow anesthesia.
- To evaluate desflurane use in Flow-I and Aisys anesthesia machines.
- To compare the efficacy of two different automated gas delivery devices.
Main Methods:
- 160 patients undergoing robot-assisted laparoscopic surgery were divided into four groups.
- Groups included Flow-I with automated or conventional low-flow (1 L/min) and Aisys with end-tidal gas control or conventional low-flow.
- Anesthesia was maintained at a minimum alveolar concentration (MAC) of 0.7-0.8, with desflurane consumption recorded at 9, 30, and 60 minutes.
Main Results:
- Automated systems reduced desflurane consumption by 60 minutes: Aisys from 25.8 to 15.2 mL (P < .001) and Flow-I from 22.1 to 16.8 mL (P < .001) compared to conventional low flow.
- Aisys end-tidal control achieved target MAC and stable FGF faster than Flow-I automated gas control.
- Both automated gas delivery devices demonstrated reliability in clinical use.
Conclusions:
- Automated gas delivery devices significantly decrease desflurane consumption compared to conventional low-flow anesthesia.
- The study confirms the effectiveness and reliability of automated systems in reducing anesthetic agent usage.
- Automated anesthesia delivery systems offer a viable method for optimizing volatile anesthetic management.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Inhalational Anesthetics: Overview
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Stages of General Anesthesia
Parenteral Anesthetics: Overview