Reliability of 3 Urine Specific Gravity Meters for Measuring Brix and Urine Solutions at Different Temperatures
Floris Wardenaar1, Carmen P Ortega-Santos1, Kaila Vento1
1College of Health Solutions, Arizona State University, Phoenix.
Context:
Urine specific gravity (USG) should be measured at room temperature (20°C), but the temperature of the sample is not always considered.
Objective:
To evaluate the effect of sample temperature on the measurement accuracy of a digital refractometer (DIG), manual optical refractometer (MAN), and hydrometer (HYD).
Design:
Descriptive laboratory study.
Main Outcome Measure(S):
Urine specific gravity.
Results:
Experiment 1 (24 Brix (°Bx) samples) showed that measurements via the DIG and MAN did not differ from the reference, but HYD provided lower or inconsistent values compared with °Bx and was highly correlated with °Bx solutions (r, > = 0.89). The overall diagnostic ability of elevated USG cut-off values (≥1.020, ≥1.025, ≥1.030) was high for all tools (area under the curve >0.92). Misclassification of samples increased from 0 to 2 at 1.020 to 1 to 3 samples at cutoffs of 1.025 and 1.030 USG. Bland-Altman analysis showed that the DIG 5°C underreported slightly without reporting bias (r = -0.344, P = .13); all other plots for the DIG, MAN, and HYD showed considerably larger underreporting at higher concentrations (r = -0.21 to -0.97 with P >.02) at all temperatures. The outcomes of experiment 2 (33 fresh urine samples) using DIG 20°C as the standard demonstrated only negligible differences between the DIG and MAN at all temperatures but larger differences using the HYD.
Conclusions:
All tools showed reporting bias compared with the °Bx solutions, which can affect the classification of low and high urine concentration at higher USG cutoff values, especially with a sample temperature of 37°C.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Urine Studies I: Urinalysis
Uncertainty in Measurement: Reading Instruments
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Thermometers and Temperature Scales
As many physical properties depend on temperature, the variety of thermometers is...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...