Optimizing Identification of Power Injectable Ports on the Scout Images for Multidetector Computed Tomography
Fides R Schwartz1, Kevin S Iliffe1, Andrew Somers1
1From the Department of Radiology, Duke University Medical Center.
Objective:
The objective of this study was to assess the impact of tube voltage and image display on the identification of power ports features on anterior-posterior scout images to inform optimal workflow for multidetector computed tomography (MDCT) examinations.
Materials And Methods:
Four ports, representing variable material composition (titanium/silicone), shapes, and computed tomography (CT) markings, were imaged on an adult anthropomorphic chest phantom using a dual-source MDCT at variable peak tube voltages (80, 100, 120, 150, and Sn150 kVp). Images were reviewed at variable image display setting by 5 blinded readers to assess port features of material composition, shape, and text markings as well as overall preferred image quality.
Results:
Material composition was correctly identified for all ports by all readers across all kilovoltage-peak settings. The identification by shape was more reliable than CT markers for all but one of the ports. CT marker identification was up to 80% for titanium ports at window level settings optimized for metal (window width, 200; window center, -150) and at a soft tissue setting (window width, 400; window center, 40) for silicone ports. Interreader agreement for best image quality per kilovoltage-peak setting was moderate to substantial for 3 ports (k = 0.5-0.62) but only fair for 1 port (k = 0.27). The highest overall rank for image quality was given unanimously to Sn150 kVp for imaging titanium ports and 100 kVp for silicone ports.
Conclusions:
Power port identification on MDCT scout images can be optimized with modification of MDCT scout acquisition and display settings based on the main port material.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET