Related Experiment Video
Updated: Nov 10, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Assessment of the Exposure to Gradient Magnetic Fields Generated by MRI Tomographs: Measurement Method, Verification
Riccardo Di Liberto1, Daniele Andreuccetti2, Moreno Comelli2
1IRCCS, Scientific Hospitalization and Care Institute, Policlinico San Matteo, Viale Camillo Golgi 19, 27100 Pavia, Italy.
Abstract:
This work is the result of a campaign of measures of exposure levels to magnetic field gradients (GMF) generated by magnetic resonance imaging (MRI) tomographs, to which both healthcare staff and any persons accompanying patients who remain inside the magnet room are exposed while performing a diagnostic Investigation. The study was conducted on three MRI tomographs with a static magnetic induction field up to 1.5 T installed in two hospitals of Lombardy. The study aims to characterize electromagnetic emissions within the magnet room and the definition of a measurement method suitable for assessing the level of exposure of healthcare personnel and any persons accompanying patients. The measurements performed concerned the determination of the weighted peak index for magnetic induction, due to the diagnostic GMF, relating to the action levels for the workers and the reference levels for the general population, in force in the European Union. Thanks to the defined experimental setup, the use of two different measuring instruments, and the software resources of the WEBNIR platform, it was possible to identify, for both categories of exposed persons, the "clearance" space, i.e., the distance from the magnet of the tomograph that guarantees health protection concerning the exposure to GMF, according to the indications of the standards in force. The method used showed that the exposure levels to GMF are substantially safe for professionally exposed workers who do not carry specific risks. For workers particularly sensitive to the specific risk, as well as to individuals part of the population, it is however advisable to maintain a distance from the magnet of about one meter to prevent sensorial neuromuscular stimulation effects.
Insights
Exposure to magnetic field gradients (GMF) from MRI scans is safe for most healthcare workers. A one-meter distance is recommended for sensitive individuals and the general population to prevent stimulation effects.
Area of Science:
- Medical Physics
- Occupational Health
- Electromagnetic Fields
Background:
- Healthcare staff and accompanying persons are exposed to magnetic field gradients (GMF) during Magnetic Resonance Imaging (MRI) procedures.
- Understanding exposure levels is crucial for ensuring safety within MRI environments.
Purpose of the Study:
- To characterize electromagnetic emissions from MRI tomographs.
- To define a measurement method for assessing GMF exposure.
- To establish safe distances (clearance) within MRI rooms.
Main Methods:
- Measurements were conducted on three MRI tomographs (up to 1.5 T) in two hospitals.
- The weighted peak index for magnetic induction was determined.
- The WEBNIR platform and two distinct measuring instruments were utilized.
Main Results:
- Exposure levels to GMF were found to be substantially safe for most healthcare workers.
- A 'clearance' space was identified, ensuring health protection based on EU standards.
- For sensitive individuals and the general population, a distance of approximately one meter is advised.
Conclusions:
- The defined measurement method effectively assesses GMF exposure from MRI.
- Current exposure levels pose minimal risk to most healthcare professionals.
- Maintaining a one-meter distance is recommended for sensitive individuals to prevent potential stimulation effects.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging