Electrochemical Immunosensor for the Quantification of S100B at Clinically Relevant Levels Using a Cysteamine
Alexander Rodríguez1, Francisco Burgos-Flórez1,2, José D Posada3
1Biotechnology Research Group, Universidad del Norte, Barranquilla 081007, Colombia.
This study presents a novel biosensor for rapid S100B biomarker detection, crucial for assessing traumatic brain injury (TBI) severity. The developed interdigitated gold electrode (AuIDE) biosensor offers a faster, more cost-effective alternative to traditional assays.
Area of Science:
- Biomedical Engineering
- Biosensor Technology
- Neuroscience Diagnostics
Background:
- Traumatic brain injury (TBI) necessitates timely clinical assessment via biomarkers like S100B.
- Current S100B quantification methods (immunoassays) are slow, expensive, and require expertise.
- Development of rapid, accessible diagnostic tools for TBI management is critical.
Purpose of the Study:
- To develop and characterize a novel electrochemical biosensor for S100B quantification.
- To immobilize anti-S100B antibodies onto gold electrodes using a zero-length cross-linking approach.
- To evaluate the performance of planar (AuEs) and interdigitated (AuIDEs) gold electrode platforms.
Main Methods:
- Antibody immobilization on gold electrodes (AuEs and AuIDEs) via cysteamine SAM and carbonyl-bond cross-linking.
- Surface characterization using Atomic Force Microscopy (AFM) and Fourier-Transform Infrared Spectroscopy (FTIR).
- Electrochemical Impedance Spectroscopy (EIS) to measure changes in charge-transfer resistance (Rct) with varying S100B concentrations (10-1000 pg/mL).
Main Results:
- Increased Rct values correlated with higher S100B concentrations on both AuEs and AuIDEs.
- Achieved Limit-of-Detection (LOD) of 18 pg/mL for AuEs and 6 pg/mL for AuIDEs.
- AuIDEs demonstrated simpler fabrication, reduced time/cost, and potential for single-frequency capacitance analysis.
Conclusions:
- The developed electrochemical biosensor, particularly the AuIDE platform, offers a promising rapid detection method for S100B.
- This technology could significantly improve the timely diagnosis and management of TBI.
- The AuIDE biosensor presents a cost-effective and efficient alternative to conventional S100B immunoassays.