Development of a time of flight spectrometer based on position sensitive multi-wire proportional counters for fission
Akhil Jhingan1, N Saneesh1, M Kumar1
1Inter University Accelerator Centre, P.O. Box 10502, New Delhi 110067, India.
Abstract:
Characteristics and performance of a time of flight (TOF) spectrometer developed for performing fission mass distribution studies are presented. The spectrometer contains two TOF arms based on multi-wire proportional counters (MWPCs). Each arm has two MWPCs to form a start-stop detection system for TOF measurements. The start detector has an active area of 4 × 4 cm2. The stop detector is a two-dimensional position sensitive MWPC with an active area of 16 × 11 cm2. Salient features of the MWPCs are the use of reduced sub-millimeter wire pitches of 0.635 and 0.317 mm in the electrodes along with the use of gold plated tungsten wires of diameters 10 and 20 µm. A delay line for position electrodes is prepared using chip inductors and capacitors. Ten different configurations of MWPC were investigated for the start detector, which involved the use of three and four electrode geometries, use of different wire pitches, and use of aluminized mylar for timing electrodes. Performance results close to micro-channel plate detectors have been observed with some designs of MWPC, displaying rise times better than 2 ns with an estimated inherent time resolution of ∼100 ps FWHM. A position resolution of ∼1 mm (FWHM) has been observed. Design features of the MWPCs and their test performance results are described in this article.
More Related Videos
08:51Coulomb Explosion Imaging as a Tool to Distinguish Between Stereoisomers
Published on: August 18, 2017
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Mass Analyzers: Common Types
Mass Spectrometers
Atomic Emission Spectroscopy: Instrumentation
Mass Analyzers: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....