Related Experiment Video
Updated: Nov 10, 2025

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Design of multilayer-based diagnostics for measurement of high energy x rays and gamma rays
C Burcklen1, J von der Linden1, A Do1
1Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
Abstract:
We investigate several possible multilayer-based optic designs for future hard x-ray and gamma ray diagnostics, including the detection and measurement of the positron annihilation radiation at 511 keV. The focus is set on increasing the photon efficiency and signal-to-noise ratio, compared to a previous multilayer-based system that was successfully employed to measure spectra in the 55 keV-100 keV range. Several possible designs using multilayer coatings are discussed, including mirror-based optics and multilayer Laue lenses.