Metacarpal shaft fixation: a biomechanical comparison of dorsal plating, lag screws, and headless compression screws
Felix G E Dyrna1, Daniel M Avery2, Ryu Yoshida3
1Department of Trauma, Hand and Reconstructive Surgery, University Hospital Münster, Albert-Schweitzer-Campus 1, Building W1, 48149, Münster, Germany. felix.dyrna@icloud.com.
Background:
Metacarpal shaft fractures are common and can be treated nonoperatively. Shortening, angulation, and rotational deformity are indications for surgical treatment. Various forms of treatment with advantages and disadvantages have been documented. The purpose of the study was to determine the stability of fracture fixation with intramedullary headless compression screws in two types of metacarpal shaft fractures and compare them to other common forms of rigid fixation: dorsal plating and lag screw fixation. It was hypothesized that headless compression screws would demonstrate a biomechanical stronger construct.
Methods:
Five matched paired hands (age 60.9 ± 4.6 years), utilizing non-thumb metacarpals, were used for comparative fixation in two fracture types created by an osteotomy. In transverse diaphyseal fractures, fixation by headless compression screws (n = 7) and plating (n = 8) were compared. In long oblique diaphyseal fractures, headless compression screws (n = 8) were compared with plating (n = 8) and lag screws (n = 7). Testing was performed using an MTS frame producing an apex dorsal, three point bending force. Peak load to failure and stiffness were calculated from the load-displacement curve generated.
Results:
For transverse fractures, headless compression screws had a significantly higher stiffness and peak load to failure, means 249.4 N/mm and 584.8 N, than plates, means 129.02 N/mm and 303.9 N (both p < 0.001). For long oblique fractures, stiffness and peak load to failure for headless compression screws were means 209 N/mm and 758.4 N, for plates 258.7 N/mm and 518.5 N, and for lag screws 172.18 N/mm and 234.11 N. There was significance in peak load to failure for headless compression screws vs plates (p = 0.023), headless compression screws vs lag screws (p < 0.001), and plates vs lag screws (p = 0.009). There was no significant difference in stiffness between groups.
Conclusion:
Intramedullary fixation of diaphyseal metacarpal fractures with a headless compression screw provides excellent biomechanical stability. Coupled with lower risks for adverse effects, headless compression screws may be a preferable option for those requiring rapid return to sport or work.
Level Of Evidence:
Basic Science Study, Biomechanics.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Screw: Problem Solving
To evaluate the...
Wood Fasteners
Nails are the most common fasteners, consisting of sharp-pointed metal pins that are driven into wood using a hammer or a mechanical nail gun. They come in a variety of diameters, heads, and lengths. Nailing techniques include...