Dual-layer dual-energy CT for characterization of adrenal nodules: can virtual unenhanced images replace true
Jinjin Cao1, Simon Lennartz1,2, Anushri Parakh1
1Department of Radiology, Harvard Medical School, Massachusetts General Hospital, 55 Fruit Street, White 270, Boston, MA, 02114-2696, USA.
Purpose:
To investigate the diagnostic performance of dual-layer dual-energy CT (dlDECT) in the evaluation of adrenal nodules.
Methods:
In this retrospective study, 66 patients with triphasic dlDECT (unenhanced, venous phase (VP), delayed phase (DP)) for suspected adrenal lesions were included. Virtual unenhanced images (VUE) were derived from VP acquisitions. Reference diagnoses were established with true unenhanced (TUE) attenuation, absolute washout, follow-up imaging and pathological data. Attenuation for adrenal lesions and abdominal tissues was acquired on TUE, VUE, VP and DP images. VUE and TUE attenuation were compared in all included tissues. Characterization of adrenal nodules based on TUE and VUE attenuation was investigated. ROC analysis was used to determine an adjusted threshold for diagnosing lipid-rich adenomas.
Results:
Seventy-three adrenal nodules (mean size: 18.9 ± 8.9 mm) were identified in 66 patients (38 females, 28 males; age: 61 ± 13 years) including adenoma (n = 65), metastases (n = 2), pheochromocytoma (n = 3), adrenocortical carcinoma (n = 1) and myelolipoma (n = 2). Mean attenuation of all included tissues except for the abdominal aorta (p = 0.11) was significantly higher in VUE compared to TUE images, including the attenuation of adrenal nodules (20.0 ± 17.2 vs. 7.1 ± 19.8; p < 0.05). Classification of adrenal adenomas as lipid-rich based on VUE attenuation ≤ 10 HU yielded a sensitivity/specificity of 0.2/1.0, while an adjusted threshold of ≤ 22 HU yielded a sensitivity/specificity of 0.82/0.85.
Conclusion:
dlDECT-derived VUE images overestimated attenuation in adrenal nodules, resulting in low sensitivity for diagnosis of lipid-rich adenomas using the established 10 HU threshold. Based on an adjusted threshold (≤ 22 HU) a higher sensitivity was attained, yet at the expense of a lower specificity, warranting further validation.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET