Technical Performance of Continuous Pressure Insufflators Versus Traditional Insufflators in the Presence of Leaks
Silvana Perretta1,2,3, Andrea Spota2,4, Bernard Dallemagne1,2,3
1IHU-Strasbourg, Institute of Image-Guided Surgery, Strasbourg, France.
Abstract:
Background. New insufflators have been developed to manage leaks emanating from the working cavity during laparoscopic surgery. Little data have been published to show the performance of these insufflators. This study evaluates the performance of 2 modern continuous pressure insufflators in various leak conditions. Methods. Performance of a new continuous pressure insufflator, EVA15 (Palliare, Galway, Ireland), was compared with that of a continuous pressure insufflator, Airseal intelligent Flow System (iFS) (CONMED, Utica, NY), and a traditional intermittent pressure insufflator, Endoflator 40 (KARL STORZ, Tuttlingen, Germany). Six different leak scenarios were created in a swine model to create different dynamic leak profiles of between 5 and 15 L/min. Pressure deviation from setting was measured for each insufflator tested. The following were calculated: (a) mean absolute difference of measured insufflation pressure vs. the insufflator pressure setting for the 50 second measurement period (MAD) and (b) standard deviation of the absolute differences (SD). Results. The average mean absolute pressure difference between the EVA15 and Airseal is .27 mmHg. However, the average mean absolute pressure difference between the EVA15 and Endoflator 40 is 2.62 mmHg. Conclusions. Continuous pressure insufflators provide superior pressure delivery performance in leak situations when compared against a traditional intermittent pressure insufflator. No clinically significant performance difference was observed when comparing the EVA15 and Airseal iFS insufflators in the tested leak situations.
More Related Videos
04:31Reverse Needle Continuous Suture of the Pancreatic Duct to Jejunal Mucosal Pancreaticointestinal Anastomosis in Laparoscopic Pancreaticoduodenectomy
Published on: August 29, 2025
05:40Continuous Telemetric In Utero Tracheal Pressure Measurements in Fetal Lambs
Published on: December 22, 2023
Related Concept Videos
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...