Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed-mode
Thomas H Walter1, Melvin Blaze M T1, Cheryl Boissel1
1Waters Corporation, 34 Maple Street, Milford, MA, USA.
Rationale:
Mixed-mode reversed-phase/anion exchange liquid chromatography is useful for separations of mixtures containing anions (e.g. ionized acids). However, when using this form of liquid chromatography with mass spectrometry detection, the bleed of amine-containing hydrolysis products from the columns may cause ion suppression or enhancement.
Methods:
Using electrospray ionization tandem quadrupole mass spectrometry detection, we determined the ion suppression or enhancement caused by column bleed for three mixed-mode reversed-phase/weak anion-exchange columns containing stationary phases that differ in chemical structure. Two of the stationary phases are based on silica particles, while the third uses ethylene-bridged hybrid organic/inorganic particles, which have improved hydrolytic stability. Mixtures of acidic and basic analytes were combined with the chromatography flow postcolumn, both with and without a column, and their mass spectrometry ion signal responses (peak areas) were determined. The ratio of signal response with and without a column is the matrix factor. Positive ion electrospray measurements were carried out using 0.1% formic acid (pH ~ 2.7) as a mobile phase additive, and 10mM ammonium formate (pH ~ 6.4) was used for negative ion electrospray detection.
Results:
The matrix factors under both positive and negative ionization modes were closest to 1 (0.74-1.16) for the hybrid particle-based columns, showing minimal ion suppression or enhancement. In contrast, the silica-based columns gave matrix factors ranging from 0.04 to 1.86, indicating high levels of ion suppression or enhancement. These results may be explained by the differences in the structures of the stationary phases, which affect the relative amounts of hydrolysis products that elute from the columns.
Conclusions:
The low levels of mass spectrometry ion suppression or enhancement caused by column bleed from the hybrid particle-based columns should allow for accurate quantitative mass spectrometric detection combined with mixed-mode reversed-phase/weak anion-exchange chromatography.
More Related Videos
10:41Ion Exchange Chromatography IEX Coupled to Multi-angle Light Scattering MALS for Protein Separation and Characterization
Published on: April 5, 2019
07:54Capillary Electrophoresis Mass Spectrometry Approaches for Characterization of the Protein and Metabolite Corona Acquired by Nanomaterials
Published on: October 27, 2020
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...