Related Experiment Video
Updated: Nov 9, 2025

Author Spotlight: Enhancing Success of Ultrasound-Guided Neuraxial Anesthesia in Cases with Difficult Anatomy
Published on: January 31, 2025
An advanced double intravenous vasopressor automated system to treat hypotension during spinal anaesthesia for
Singaraselvan Nagarajan1, Jason Ju In Chan, Chin Wen Tan
1From the Department of Women's Anaesthesia, KK Women's and Children's Hospital (SN, JJIC, CWT, ZGAAH, AT, BLS), Duke-NUS Medical School (SN, JJIC, CWT, AT, BLS), and Centre for Quantitative Medicine, Duke-NUS Medical School, Singapore, Singapore (RS).
Background:
During spinal anaesthesia for caesarean section, haemodynamic instability may lead to maternal and foetal complications. We developed a novel advanced double intravenous vasopressor automated system (ADIVA) by using a continuous blood pressure and heart rate monitor. Treatment of hypotension was based on three criteria: the drug (phenylephrine or ephedrine) according to the heart rate; the dose of vasopressor determined by the degree of hypotension; a fast or slow bolus of vasopressor administered depending on whether there was a negative or positive gradient of SBP changes, respectively.
Objective:
The aim of this pilot study was to investigate the feasibility of the ADIVA algorithm.
Design:
A prospective pilot study.
Setting:
Single obstetrics and gynaecology centre in Singapore.
Patients:
Women undergoing elective caesarean delivery under spinal anaesthesia.
Intervention:
Automated administration of ephedrine or phenylephrine based on changes in blood pressure and heart rate (via the ADIVA algorithm) detected on continuous noninvasive haemodynamic monitoring using noninvasive continuous haemodynamic monitor (Nexfin).
Main Outcome Measures:
The primary outcome was the incidence of hypotension, defined as SBP less than 80% of baseline. The secondary outcome measures were reactive hypertension, total vasopressor requirement, maternal and neonatal outcomes and system performance.
Results:
Forty-five women were recruited. Thirty-one women (69.9%) had at least one reading of hypotension. SBP was within ±20% of the baseline in a mean ± SD of 79.7 ± 17.6% of measurements. Forty-four (97.8%) women required phenylephrine before delivery, while 15 (33.3%) required ephedrine. No rescue medications were required. Three women (6.7%) had nausea and two (4.4%) vomiting. All neonates had APGAR scores of 9 at 5 min.
Conclusion:
The ADIVA system, with noninvasive continuous haemodynamic monitoring, was able to maintain maternal SBP within ±20% of baseline for the vast majority of the measurements. This system had good maternal and foetal outcomes with minimal intervention from the attending anaesthetist.
Trial Registration:
NCT03620942.
More Related Videos
09:14Pre-Chiasmatic, Single Injection of Autologous Blood to Induce Experimental Subarachnoid Hemorrhage in a Rat Model
Published on: June 18, 2021
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
Related Concept Videos
Local Anesthetics: Clinical Application as Spinal Anesthesia
Local Anesthetics: Clinical Application as Epidural Anesthesia
Since epidural anesthetics can be infused through an epidural catheter, all types of drugs, including short-acting ones, can be administered. Chloroprocaine and lidocaine are examples of short and long-duration anesthetics, respectively. Bupivacaine...
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Cardiopulmonary Resuscitation IV: Pharmacological Management
Drug Delivery: Parenteral Route
There are three primary parenteral routes: intravenous (IV), intramuscular (IM), and subcutaneous (SC). The IV route introduces the drug directly into the bloodstream, ensuring immediate action. The IM route...
Parenteral Anesthetics: Overview