Related Experiment Video
Updated: Nov 9, 2025

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Compatibility of the Drawover Vaporizer With Military Field Ventilators
Brian E Benham1, Col Steven S Kertes2
1is a retired Army Nurse Corps officer and currently is a clinical professor teaching didactic instruction for the US Army Graduate Program in Anesthesia Nursing, Ft Sam Houston, Texas.
Abstract:
For far-forward resuscitative care for combat trauma casualties, the field anesthesia machine best suited is the Universal Portable Anesthesia Complete (UPAC, Datex Ohmeda, GE Healthcare). However, no provision was made for mechanical ventilation. Historically the Impact Uni-Vent 754 ventilator (Impact Instrumentation, Zoll Medical Corp) was used to provide hands-free ventilation but has limitations. There are 3 other field-capable ventilators that could fill the gap, but no known published research studies exist that examine the compatibility of these ventilators with the UPAC. The research question of this experimental, crossover design study was: Are there operational differences in the Impact 754, Impact Uni-Vent 731, Hamilton-T1 (Hamilton Medical Inc), and second-generation Simplified Automated Ventilator (SAVe II, AutoMedx) ventilators' performance relative to tidal volumes, rates, and inhaled anesthetic concentrations when used with the UPAC? The Impact 754 was most accurate (P<.05), followed by the Impact 731, and Hamilton-T1, and the SAVe II was the least accurate related to minute ventilation comparisons. Each ventilator demonstrated effective inhaled anesthetic delivery (r=0.97-0.99). Data analysis results demonstrated statistical and clinical significance in the reliability of the tested ventilators to deliver set volumes and rates while delivering predictable and accurate amounts of volatile anesthetic.
More Related Videos
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
10:44Comparing the Effects of Electronic Cigarette Vapor and Cigarette Smoke in a Novel In Vivo Exposure System
Published on: May 24, 2017
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Vapor Pressure Lowering