Feasibility and Safety of the Novel Vivolight Frequency Domain Optical Coherence Tomography System: A Multicenter
Qinhua Jin, Juying Qian, Yujie Zhou
1Department of Cardiology, The General Hospital of the People's Liberation Army, 28 Fuxing Road, Beijing (Wukesong), China. cyundai@vip.163.com.
Objective:
The study sought to assess the effectiveness and safety of the novel P60 Vivolight frequency-domain optical coherence tomography (OCT) system (Shenzhen Vivolight Medical Device & Technology).
Methods:
A total of 90 patients were enrolled from 3 institutions. The pullbacks were performed with both the P60 Vivolight OCT system and the Ilumien Optis OCT system (Abbott Vascular). The primary endpoint was the clear stent length (CSL). Device safety was assessed by the record of serious procedure-related or postprocedure adverse events. The secondary endpoints were the average lumen area of stent, clear image length (CIL), system stability, and imaging catheter operability.
Results:
The mean relative errors of CSL were 3.30% (95% confidence interval [CI], -0.71 to 7.31) in the full analysis set (FAS) and 0.83% (95% CI, -1.79 to 3.45) in the per-protocol set (PPS). The mean relative errors of the average lumen area of stent were 2.20% (95% CI, 0.70 to 3.80) in the FAS and 1.55% (95% CI, 0.30 to 2.80) in the PPS. No difference was observed in the percentage of obtaining >24 mm of CIL (93.18% in the P60 Vivolight group vs 95.45% in the Ilumien Optis group; P=.48). There were no serious procedure-related or postprocedure adverse events.
Conclusions:
The feasibility and safety of the novel Vivolight OCT system is equivalent to that of the Ilumien Optis OCT system.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Clinical Trials: Overview