Differential white blood cell counts in rabbits: a comparison of the Advia 2120 and a manual method

Ioannis L Oikonomidis1, Elspeth Milne1, Chiara Piccinelli1

  • 1Easter Bush Pathology, Royal (Dick) School of Veterinary Studies and The Roslin Institute, University of Edinburgh, Roslin, UK.

Insights

The Advia 2120 automated analyzer shows high correlation for rabbit heterophils and lymphocytes but underestimates heterophils and overestimates lymphocytes. Method-specific reference intervals are recommended for accurate Advia 2120 CBC results in rabbits.

Area of Science:

  • Veterinary Hematology
  • Comparative Pathology
  • Laboratory Medicine

Background:

  • Accurate differential leukocyte counts are crucial for diagnosing diseases in rabbits.
  • Automated hematology analyzers offer efficiency but require validation for each species.

Purpose of the Study:

  • To evaluate the performance of the Advia 2120 differential leukocyte count (A-Diff) against manual differential counts (M-Diff) in rabbits.
  • To identify potential biases and limitations of the Advia 2120 in rabbit hematology.

Main Methods:

  • Comparison of Advia 2120 A-Diff with M-Diff on 92 rabbit blood samples.
  • Statistical analysis included correlation (Pearson's r), Passing-Bablok regression, and Bland-Altman analysis.
  • Manual counts were performed by two blinded observers on blood smears.

Main Results:

  • High correlation (r > 0.90) for heterophils and lymphocytes.
  • Moderate correlation for monocytes (r=0.645) and high for basophils (r=0.823).
  • Low correlation for eosinophils (r=0.336).
  • Advia 2120 showed significant underestimation of heterophils and overestimation of lymphocytes.
  • Constant and proportional errors were detected for several leukocyte types.

Conclusions:

  • The Advia 2120 demonstrates variable accuracy for differential leukocyte counts in rabbits.
  • Method-specific reference intervals are necessary for Advia 2120 complete blood count (CBC) results in rabbits.
  • Manual blood smear evaluation remains essential for verifying abnormal results and cytograms.