Related Experiment Video
Updated: Nov 9, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
EFFECTIVE DOSE IN PEDIATRIC INTERVENTIONAL CARDIOLOGY
1Radiation Protection Laboratory, Saint-Petersburg Research Institute of Radiation Hygiene, Saint Petersburg, Mira st. 8, 197101 St.Petersburg, Russian Federation, Russia.
Insights
This study estimated effective doses (EDs) for pediatric interventional cardiology exams. Dose conversion coefficients (DCCs) varied significantly with patient age, with newborns receiving the highest doses.
Area of Science:
- Medical Physics
- Pediatric Radiology
- Radiation Dosimetry
Background:
- Pediatric interventional cardiology procedures involve ionizing radiation.
- Accurate estimation of effective doses (EDs) and dose conversion coefficients (DCCs) is crucial for pediatric radiation protection.
- Variability in patient parameters and procedure techniques can influence radiation doses.
Purpose of the Study:
- To estimate effective doses (EDs) for children undergoing interventional cardiology examinations in St. Petersburg.
- To calculate associated dose conversion coefficients (DCCs).
- To investigate the dependence of DCCs on patient age and examination parameters.
Main Methods:
- Data collection from over 300 pediatric interventional cardiology examinations across three hospitals.
- Categorization of patients into five age groups: newborn, 1, 5, 10, and 15 years old.
- Calculation of EDs and DCCs using PCXMC 2.0 software based on collected technical, geometrical, dosimetric, and patient parameters.
Main Results:
- Significant dependence of DCCs on patient age and beam quality was observed.
- DCCs varied by a factor of 10 between the newborn and 15-year-old groups.
- The maximum estimated effective dose was 26 mSv, found in the newborn age group.
Conclusions:
- Patient age is a critical factor influencing radiation dose conversion coefficients in pediatric interventional cardiology.
- Radiation protection strategies should account for age-specific dose variations.
- Further research into optimizing pediatric interventional cardiology procedures to minimize radiation exposure is warranted.
Abstract:
The aim of this study was to estimate the effective doses (EDs) to children undergoing pediatric interventional cardiology examinations in hospitals of St.-Petersburg, to calculate associated dose conversion coefficients (DCCs) and to investigate their dependence of different parameters of the various procedures. Basic parameters of pediatric cardiology examinations and patient doses were studied in three main pediatric city hospitals. More than 300 examinations were under consideration. The technical, geometrical and dosimetric parameters, as well as patients' parameters, for each individual procedure were collected. All cardiac examinations were divided into five age groups (newborn, 1, 5, 10 and 15 y old). The EDs and the corresponding DCCs from the dose-area product to ED were calculated using PCXMC 2.0 software and based on the collected data. The study included the evaluation of the relations between the DCCs, patient age and parameters of examinations. The results of the study indicated significant dependence of the DCCs on the patients' age and the beam quality. The DCCs for the newborns and 15-year old groups differed by a factor of 10. The maximum estimated ED of 26 mSv was found in the newborn age group.
Related Concept Videos
Drug Dosing: Infants and Children
Pharmacokinetics in Pediatric Patients: Drug Excretion
Pharmacokinetics in Pediatric Patients: Drug Distribution
Dose Size and Dosing Frequency: Determination Methods
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Drug Dosing: Geriatric Patients