Related Experiment Video
Updated: Nov 9, 2025

Total Internal Reflection Absorption Spectroscopy TIRAS for the Detection of Solvated Electrons at a Plasma-liquid Interface
Published on: January 24, 2018
Hollow electrode capillary plasma ionization source for rapid online detection of gaseous samples
Wen Xu1,2, Zeyue Shen1,2, Junliang Zhang1,2
1Faculty of Electrical Engineering and Computer Science, Ningbo University, Ningbo, China.
Rationale:
Gas chromatography mass spectrometry (GC-MS) with electron ionization (EI) is the most widely used analysis technique of gaseous samples, but it may be time-consuming for online monitoring of mixtures whose concentrations relatively change rapidly. On the contrary, current ionization methods, such as chemical ionization (CI) and proton transfer reaction (PTR), also have some disadvantages such as selectivity. Therefore, appropriate soft ionization sources are searched for rapid online detection.
Methods:
Hollow electrode capillary plasma ionization (HECPI) is based on single electrode plasma. A hollow capillary was placed as both the electrode and the inlet of the gaseous samples. The ionization source is coupled with a mass spectrometer for performance evaluation.
Results:
Several typical compounds have been tested with HECPI-mass spectrometer. In this process, the dominant ion peaks of all compounds can be indexed as molecular ion peaks, and the product ions of HECPI are less than that of dielectric barrier discharge ionization (DBDI). Three gaseous samples (linalool, triethylamine, and styrene) with various concentrations have been used to further confirm the performance of this source, and the detection limit of linalool is as low as 10 ppb.
Conclusions:
HECPI is simple in structure and shows good performance. The results also show that HECPI has the potential to be an effective tool for detecting online gaseous samples rapidly.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Gas Chromatography: Types of Detectors-II
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Capillary Electrophoresis: Instrumentation
Atomic Emission Spectroscopy: Lab