Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Validation of an algorithm for continuous monitoring of atrial fibrillation using a consumer smartwatch
Robert Avram1, Mattheus Ramsis1, Ashley D Cristal1
1Division of Cardiology, University of California San Francisco, San Francisco, California.
Background:
Consumer devices with broad reach may be useful in screening for atrial fibrillation (AF) in appropriate populations. However, currently no consumer devices are capable of continuous monitoring for AF.
Objective:
The purpose of this study was to estimate the sensitivity and specificity of a smartwatch algorithm for continuous detection of AF from sinus rhythm in a free-living setting.
Methods:
We studied a commercially available smartwatch with photoplethysmography (W-PPG) and electrocardiogram (W-ECG) capabilities. We validated a novel W-PPG algorithm combined with a W-ECG algorithm in a free-living setting, and compared the results to those of a 28-day continuous ECG patch (P-ECG).
Results:
A total of 204 participants completed the free-living study, recording 81,944 hours with both P-ECG and smartwatch measurements. We found sensitivity of 87.8% (95% confidence interval [CI] 83.6%-91.0%) and specificity of 97.4% (95% CI 97.1%-97.7%) for the W-PPG algorithm (every 5-minute classification); sensitivity of 98.9% (95% CI 98.1%-99.4%) and specificity of 99.3% (95% CI 99.1%-99.5%) for the W-ECG algorithm; and sensitivity of 96.9% (95% CI 93.7%-98.5%) and specificity of 99.3% (95% CI 98.4%-99.7%) for W-PPG triggered W-ECG with a single W-ECG required for confirmation of AF. We found a very strong correlation of W-PPG in quantifying AF burden compared to P-ECG (r = 0.98).
Conclusion:
Our findings demonstrate that a novel algorithm using a commercially available smartwatch can continuously detect AF with excellent performance and that confirmation with W-ECG further enhances specificity. In addition, our W-PPG algorithm can estimate AF burden. Further research is needed to determine whether this algorithm is useful in screening for AF in select at-risk patients.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
15:00Setup of Consumer Wearable Devices for Exposure and Health Monitoring in Population Studies
Published on: February 3, 2023
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of blood pressure in brachial artery(two-step method)