Purpose:
Hyperspectral imaging (HSI) is a noninvasive spectroscopy technique for determining superficial tissue oxygenation. The HyperView™ system is a hand-held camera that enables perfusion image acquisition. The evaluation of superficial tissue oxygenation is warranted in the evaluation of patients with peripheral arterial disease. The aim was to determine the reliability of repeated HSI measurements.
Methods:
In this prospective cohort study, HSI was performed on 50 healthy volunteers with a mean age of 26.4 ± 2.5 years, at the lower extremity. Two independent observers performed HSI during two subsequent measurement sessions. Short term test-retest reliability and intra- and inter-observer reliability were determined, and generalizability and decision studies were performed. Transcutaneous oxygen pressure (TcPO2) measurements were also performed.
Results:
The short term test-retest reliability was good for the HSI values determined at the lower extremity, ranging from 0.72 to 0.90. Intra- and inter-observer reliability determined at different days were poor to moderate for both HSI (0.24 to 0.71 and 0.30 to 0.58, respectively) and TcPO2 (0.54 and 0.56, and 0.51 and 0.31, respectively). Reliability can be increased to >0.75 by averaging two measurements on different days.
Conclusion:
This study showed good short term test-retest reliability for HSI measurements, however low intra- and inter-observer reliability was observed for tissue oxygenation measurements with both HSI and TcPO2 performed at separate days in young healthy volunteers. Reliability of HSI can be improved when determined as a mean of two measurements taken on different days.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessing Blood pressure in the Leg
Preparation: