Risk of Cardiac Implantable Electronic Device Malfunctioning During Pencil Beam Proton Scanning in an In Vitro
Henrik Laurits Bjerre1, Mads Brix Kronborg2, Jens Cosedis Nielsen2
1Department of Cardiology, Aarhus University Hospital, Denmark; Danish Centre for Particle Therapy, Aarhus University Hospital, Denmark.
Purpose:
Cardiac implantable electronic devices (CIED) are sensitive to scattered secondary neutrons from proton beam irradiation. This experimental in vitro study investigated risk of CIED errors during pencil beam proton therapy.
Methods And Materials:
We used 62 explanted CIEDs from 4 manufacturers; 49 CIEDs were subjected to a simulated clinical protocol with daily 2 Gy relative biological effectiveness fractions prescribed to the phantom. Devices were located at 3 different lateral distances from the spread-out Bragg peak to investigate the risk of permanent or temporary device errors. Additionally, 13 devices with leads connected were monitored live during consecutive irradiations to investigate the risk of noise, over- or undersense, pace inhibition, and inappropriate shock therapy.
Results:
We detected 61 reset errors in 1728 fractions, and all except 1 CIED were reprogrammed to normal function. All, except 1 reset, occurred in devices from the same manufacturer. These were successfully reprogrammed to normal function. The 1 remaining CIED was locked in permanent safety mode. Secondary neutron dose, as estimated by Monte Carlo simulations, was found to significantly increase the odds of CIED resets by 55% per mSv. Clinically significant battery depletion was observed in 5 devices. We observed no noise, over- or undersense, pace inhibition, or inappropriate shock therapy during 362 fractions of live monitoring.
Conclusions:
Reprogrammable CIED reset was the most commonly observed malfunction during proton therapy, and reset risk depended on secondary neutron exposure. The benefits of proton therapy are expected to outweigh the risk of CIED malfunctioning for most patients.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...