Andrea Padoan1,2, Francesco Bonfante3, Chiara Cosma2
1Department of Medicine-DIMED, Medical School, University of Padova, Padova, Italy.
Objectives:
SARS-CoV-2 serology presents an important role in several aspects of COVID-19 pandemic. Immunoassays performances have to be accurately evaluated and correlated with neutralizing antibodies. We investigated the analytical and clinical performances of a SARS-CoV-2 RBD IgG assay, automated on a high throughput platform, and the correlation of the antibodies (Ab) levels with the plaque reduction neutralization (PRNT50) Ab titers.
Methods:
A series of 546 samples were evaluated by SARS-CoV-2 RBD IgG assay (Snibe diagnostics), including 171 negative and 168 positive SARS-CoV-2 subjects and a further group of 207 subjects of the COVID-19 family clusters follow-up cohort.
Results:
Assay imprecision ranged from 3.98 to 12.18% being satisfactory at low and medium levels; linearity was excellent in all the measurement range. Considering specimens collected after 14 days post symptoms onset, overall sensitivity and specificity were 99.0 and 92.5%, respectively. A total of 281 leftover samples results of the PRNT50 test were available. An elevated correlation was obtained between the SARS-CoV-2 RBD IgG assay and the PRNT50 titer at univariate (ρ=0.689) and multivariate (ρ=0.712) analyses.
Conclusions: