Validation of a new wearable device for type 3 sleep test without flowmeter
Mauro Contini1, Antonio Sarmento2, Paola Gugliandolo1
1Centro Cardiologico Monzino, IRCCS, Milano, Italy.
Background:
Ventilation monitoring during sleep is performed by sleep test instrumentation that is uncomfortable for the patients due to the presence of the flowmeter. The objective of this study was to evaluate if an innovative type 3 wearable system, the X10X and X10Y, is able to correctly detect events of apnea and hypopnea and to classify the severity of sleep apnea without the use of a flowmeter.
Methods:
40 patients with sleep disordered breathing were analyzed by continuous and simultaneous recording of X10X and X10Y and another certified type 3 system, SOMNOtouch, used for comparison. Evaluation was performed in terms of quality of respiratory signals (scores from 1, lowest, to 5, highest), duration and classification of apneas, as well as identification and duration of hypopneas.
Results:
580 periods were evaluated. Mean quality assigned score was 3.37±1.42 and 3.25±1.35 for X10X and X10Y and SOMNOtouch, respectively. The agreement between the two systems was evaluated with grades 4 and 5 in 383 out of 580 cases. A high correlation (r2 = 0.921; p<0.001) was found between the AHI indexes obtained from the two systems. X10X and X10Y devices were able to correctly classify 72.3% of the obstructive apneas, 81% of the central apneas, 61.3% of the hypopneas, and 64.6% of the mixed apneas when compared to SOMNOtouch device.
Conclusion:
The X10X and X10Y devices are able to provide a correct grading of sleep respiratory disorders without the need of a nasal cannula for respiratory flow measurement and can be considered as a type 3 sleep test device for screening tests.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...