Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
A new phantom developed to test the ATCM performance of chest CT scanners
Yang Yang1, Weihai Zhuo1, Bo Chen1
1Institute of Radiation Medicine, Fudan University, Shanghai 200032,People's Republic of China.
Abstract:
In this study, a new ATCM phantom was developed to test the performance of the automatic tube current modulation (ATCM) of computed tomography (CT) scanners.. Based on the Chinese reference man and Monte Carlo simulations of x-ray attenuation, a more realistic ATCM phantom made of polymethyl methacrylate was developed. The phantom has a length of 20 cm, and it can be used to measure the dose profile along the central axis using 19 real-time MOSFET detectors. The image noise can be calculated slice by slice in the phantom's center. Test experiments showed that the phantom could initiate tube current modulation under different modulation levels of CT scans, and the actual effects of ATCM could be evaluated with the aid of the dose profile measurements. Using the measured dose profiles and image noise, the preferred dose can easily be identified from a choice of different modulation levels. The new phantom developed in this study can be used to test the ATCM performance of CT scanners, and is useful for further studies of the optimization of CT scan protocols with ATCM.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT