Procurement of Parathyroid Glands from Living Donor Pigs and Ex Vivo Identification
Published on: August 1, 2025
Initial clinical experiences using the intraoperative probe-based parathyroid autofluorescence identification
Colleen M Kiernan1, Giju Thomas2,3, Naira Baregamian1
1Division of Surgical Oncology and Endocrine Surgery, Department of Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Background And Objective:
The Food and Drug Administration has cleared a probe-based near-infrared autofluorescence (NIRAF) detection system called PTeye™ as an adjunct tool for label-free intraoperative parathyroid gland (PG) identification. Since PTeye™ has been investigated only in a "blinded" manner to date, this study describes the preliminary impressions of PTeye™ when used by surgeons without being blinded to the device output.
Methods:
Patients undergoing thyroid and parathyroid procedures were prospectively recruited. Target tissues were intraoperatively assessed with PTeye™. The surgeon's confidence in PG identification was recorded concomitantly with NIRAF parameters that were output in real-time from PTeye™.
Results:
A retrospective review of prospectively collected data on 83 patients was performed. PTeye™ was used for interrogating 336 target tissues in 46 parathyroid and 37 thyroid procedures. PTeye™ yielded an overall accuracy of 94.3% with a positive predictive value of 93.0% and a negative predictive value of 100%. An increase in confidence for intraoperative PG identification with PTeye™ was observed by all three participating high-volume surgeons, irrespective of their level of accrued surgical experience.
Conclusions:
Probe-based NIRAF detection with PTeye™ can be a valuable adjunct device to intraoperatively identify PGs for surgeons of varied training and experience.