Short-term Evaluation of Negative Pressure Applied by the Multi-Pressure Dial System to Lower Nocturnal IOP: A
Jeffrey L Goldberg1, Jesus Jiminez-Roman2, Alejandra Hernandez-Oteyza3
1Byers Eye Institute, Stanford University, Stanford, CA, USA.
Introduction:
The purpose of this study was to investigate the short-term safety and feasibility of negative pressure application by the Multi-Pressure Dial (MPD) System to lower nocturnal intraocular pressure (IOP) in subjects with open-angle glaucoma (OAG).
Methods:
A prospective, controlled, intra-subject study of 22 eyes from 11 subjects at a single site was performed. All subjects had a history of OAG and were currently using a topical prostaglandin. For each subject, the eye with the highest IOP in the supine position was selected as the treatment eye (TE) and the contralateral eye served as the control eye (CE). The negative pressure for the TE was set to 60% of the baseline IOP value with no negative pressure in the CE. IOP measurements were collected at three prespecified time points overnight in the supine position with active negative pressure. The primary outcome measure was mean IOP with the application of negative pressure.
Results:
At the three overnight time points, the mean (± standard deviation) baseline IOP prior to negative pressure application was 22.2 ± 2.5 mmHg in the TE and 21.8 ± 2.5 mmHg in the CE. With the application of 60% negative pressure to the TE and no active negative pressure to the CE, the mean IOP was 14.2 ± 2.2 and 19.5 ± 2.4 mmHg, respectively. The mean percentage IOP reduction in the TE was 35% (p < 0.001). There were two minor adverse events, both unrelated to device wear, and there were no IOP spikes ≥ 10 mmHg.
Conclusion:
The MPD can safely and effectively lower nocturnal IOP in the supine position. The MPD holds promise as a potential new, non-invasive treatment option for the control of nocturnal IOP.