Detection and removal of pacing artifacts prior to automated analysis of 12-lead ECG
Kazi T Haq1, Neeraj Javadekar1, Larisa G Tereshchenko1
1Oregon Health & Science University, Knight Cardiovascular Institute, Portland, OR, USA.
Background:
Pacing artifacts must be excluded from the analysis of paced ECG waveform. This study aimed to develop and validate an algorithm to identify and remove the pacing artifacts on ECG and vectorcardiogram (VCG).
Methods:
We developed a semi-automatic algorithm that identifies the onset and offset of a pacing artifact based on the VCG signal slope steepness and designed a graphical user interface that permits quality control and fine-tuning the constraining threshold values. We used 1054 ECGs from the retrospective, multicenter cohort study "Global Electrical Heterogeneity and Clinical Outcomes," including 3825 atrial and 10,031 ventricular pacing artifacts for the algorithm development and 22 ECGs including 108 atrial and 241 ventricular pacing artifacts for validation. Validation was performed per digital sample. We used the kappa-statistic of interrater agreement between manually labeled sample (ground-truth) and automated detection.
Results:
The constraining parameter values were for onset threshold 13.06 ± 6.21 microV/ms, offset threshold 34.77 ± 17.80 microV/ms, and maximum window size 27.23 ± 3.53 ms. The automated algorithm detected a digital sample belonging to pacing artifact with a sensitivity of 74.5% and specificity of 99.6% and classified correctly 98.8% of digital samples (ROC AUC 0.871; 95%CI 0.853-0.878). The kappa-statistic was 0.785, indicating substantial agreement. The agreement was on 98.81% digital samples, significantly (P < 0.00001) larger than the random agreement on 94.43% of digital samples.
Conclusions:
The semi-automated algorithm can detect and remove ECG pacing artifacts with high accuracy and provide a user-friendly interface for quality control.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias