Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Variation of heart and lung radiation doses according to setup uncertainty in left breast cancer
Sunmin Park1, Chai Hong Rim1, Won Sup Yoon2
1Department of Radiation Oncology, Ansan Hospital, Korea University, 123 Jeokgeum-ro, Danwon-gu, Ansan, Gyeonggi-do, 15355, Republic of Korea.
Purpose:
Breast radiotherapy set-up is often uncertain. Actual dose distribution to normal tissues could be different from planned dose distribution. The objective of this study was to investigate such difference in dose distribution according to the extent of set-up error in breast radiotherapy.
Materials And Methods:
A total of 50 Gy with fraction size of 2 Gy was given to 30 left breasts with different set-ups applying a deep inspiration breath holding (DIBH) or a free breathing (FB) technique. Under the assumption that errors might come from translational axes of deep or caudal directions, the isocenter was shifted from the original tangential alignment every 2.5 mm to simulate uncertainty of deep and caudal tangential set-up in DIBH and FB. Changes were evaluated for dosimetric parameters for the heart, the left ventricle (LV), the left anterior descending coronary artery (LAD), and the ipsilateral lung.
Results:
On the original plan, mean doses of heart and ipsilateral lung were 2.0 ± 1.1 Gy and 3.7 ± 1.4 Gy in DIBH and 8.4 ± 1.3 Gy and 7.8 ± 1.5 Gy in FB, respectively. The change of dose distribution for the heart in DIBH was milder than that in FB. The deeper the tangential set-up, the worse the heart, LV, LAD, and ipsilateral lung doses, showing as much as 49.4%, 56.4%, 90.3%, and 26.1% shifts, respectively, in 5 mm DIBH setup. The caudal set-up did not show significant dose difference. In multiple comparison of DIBH, differences of mean dose occurred in all 7.5 mm deep set-ups for the heart (p = 0.025), the LV (p = 0.049), and LAD (p = 0.025) in DIBH.
Conclusions:
To correct set-up error over indicated limitation for deep tangential set-up in DIBH at 5 mm action level, mean heart and ipsilateral lung doses are expected to increase approximately 50% and 25%, respectively.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...