Original Article--Outcomes of Pacing in Egyptian Pediatric Population

Ahmed Nabil Ali1, Samir S Wafa1, Hosni Hosni Arafa1

  • 1Department of Cardiology, Faculty of Medicine, Ain Shams University, Cairo, Egypt.

Insights

Pediatric pacemakers are generally safe, but lead-related complications remain a concern. Endocardial pacing demonstrated superior longevity for both batteries and leads compared to epicardial pacing in children.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering
  • Cardiac Electrophysiology

Background:

  • Permanent pacemakers are crucial for managing rhythm disturbances in pediatric patients, often necessitated by congenital heart conditions.
  • Managing pacemakers in children presents unique challenges including growth, activity levels, and infection risk.
  • Device management requires a highly individualized approach due to the complexity of congenital heart diseases.

Purpose of the Study:

  • To evaluate the long-term outcomes of permanent pacemaker implantation in a pediatric and adolescent population.
  • To compare the performance and complication rates of epicardial versus endocardial pacing systems in children.
  • To present institutional experience with pediatric pacemaker management.

Main Methods:

  • A cross-sectional observational study of 100 pediatric patients with permanent pacemakers.
  • Data collection included history, clinical examination, ECG, echocardiography, and device programming.
  • Comparison of device types, component longevity, and complications between epicardial and endocardial pacing.

Main Results:

  • Endocardial pacing (74%) was more common than epicardial pacing (26%).
  • Endocardial pacing showed significantly higher first battery longevity (108 vs. 60 months) and lead longevity (105 vs. 58 months).
  • Overall complication rate was 25%, with lead-related issues being most common (16 patients), particularly lead fracture in the endocardial group.

Conclusions:

  • Pacemakers in children are generally safe, though lead-related complications are prevalent.
  • Epicardial leads had a higher incidence of lead failure to capture compared to endocardial leads.
  • Endocardial pacemakers offer superior longevity for both batteries and leads in the pediatric population.
Abstract