Effect of different thresholds for CT perfusion volumetric analysis on estimated ischemic core and penumbral volumes
Simo Karhi1,2, Olli Tähtinen1,2, Joona Aherto1,2
1Department of Clinical Radiology, Kuopio University Hospital, Kuopio, Finland.
Purpose:
This single-center study compared three threshold settings for automated analysis of the ischemic core (IC) and penumbral volumes using computed tomographic perfusion, and their accuracy for predicting final infarct volume (FIV) in patients with anterior circulation acute ischemic stroke (AIS).
Methods:
Fifty-two consecutive AIS patients undergoing mechanical thrombectomy (November 2015-March 2018) were included. Perfusion images were retrospectively analyzed using a single CT Neuro perfusion application (syngo.via 4.1, Siemens Healthcare GmbH). Three threshold values (S1-S3) were derived from another commercial package (RAPID; iSchema View) (S1), up-to-date syngo.via default values (S2), and adapted values for syngo.via from a reference study (S3). The results were compared with FIV determined by non-contrast CT.
Results:
The median IC volume (mL) was 24.6 (interquartile range: 13.7-58.1) with S1 and 30.1 (20.1-53.1) with S2/S3. After removing the contralateral hemisphere from the analysis, the median IC volume decreased by 1.33(0-3.14) with S1 versus 9.13 (6.24-14.82) with S2/S3. The median penumbral volume (mL) was 74.52 (49.64-131.91), 77.86 (46.56-99.23), and 173.23 (125.86-200.64) for S1, S2, and S3, respectively. Limiting analysis to the affected hemisphere, the penumbral volume decreased by 1.6 (0.13-9.02), 19.29 (12.59-26.52), and 58.33 mL (45.53-74.84) for S1, S2, and S3, respectively. The correlation between IC and FIV was highest in patients with successful recanalization (n = 34, r = 0.784 for S1; r = 0.797 for S2/S3).
Conclusion:
Optimizing thresholds significantly improves the accuracy of estimated IC and penumbral volumes. Current recommended values produce diversified results. International guidelines based on larger multicenter studies should be established to support the standardization of volumetric analysis in clinical decision-making.
More Related Videos
09:48Quantification of Neurovascular Protection Following Repetitive Hypoxic Preconditioning and Transient Middle Cerebral Artery Occlusion in Mice
Published on: May 4, 2015
10:31Non-invasive Imaging and Analysis of Cerebral Ischemia in Living Rats Using Positron Emission Tomography with 18F-FDG
Published on: December 28, 2014
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Imaging Studies for Cardiovascular System V: CT